Surgical Case Cart Washer Utensil Washer
Surgical Case Cart Washer Specifications

Surgical Case Cart Washer 480LX
Surgical Case Cart Washer 486LX

SPECIFICATIONS

Surgical Case Cart Washer Utensil Washer APPLICATION
The Cart & Utensil Washer shall be a heavy duty, large capacity, hydro-spray washer/dryer designed for use in dietary and central processing departments to clean, sanitize and dry surgical cases, general purpose carts, dietary carts, racked utensils and utility items used in health care facilities.  Pit mounting is standard.

Surgical Case Cart Washer Utensil Washer Size:
480 LX Compartment40" wide x 85"   high x 170" long
480 LX Overall Unit*101" wide x 121" high x 205" long

486 LX Compartment- 46" wide x 85" high x 170" long
486 LX Overall Unit*- 107" wide x 121" high x 205" long

*Surgical Case Cart Washer Utensil Washer Size may vary depending upon options chosen.

Surgical Case Cart Washer Utensil Washer
STANDARD FEATURES

Surgical Case Cart Washer Utensil Washer AUTOMATIC SELF-CLEANING DEBRIS SCREEN
The treatment pump shall be provided with a self-cleaning debris screen having perforations considerably smaller machined jet orifices to prevent plugged jets.  The screen shall be inter-piped and inter-wired with the control system to filter all re-circulated solutions and automatically back flush debris to the building drain.

Surgical Case Cart Washer Utensil Washer MODULAR CONSTRUCTION
Washer shall be designed for disassembly into sections for entry into existing building.

Surgical Case Cart Washer Utensil Washer OSCILLATING SPRAY JET SYSTEM
The washer shall be provided with an oscillating jet system for all treatment solutions.  The system shall consist of machined jets mounted on spray trees suspended from an oscillating carriage traveling on self-lubricating machined Delrin? wheels. The jets are positioned to reach all cart surfaces including the underside of shelves and base.  The system is driven with a bi-directional, air drive unit in conjunction with the treatment schedule and equipped with a safety system to prevent the oscillating header from damaging items being washed.  Interference of greater than 15 seconds will be annunciated visually and audibly  to the operator. 

Surgical Case Cart Washer Utensil Washer TEMPERATURE ASSURANCE
The operator may select to assure the temperatures for each phase of the cycle.  If the temperature assurance is selected, the phase shall not begin timing until the re-circulated water temperature reaches the set point. If the re-circulated temperature drops below the set point during phase, the timer shall stop.  The phase timer resets or continues from where it stopped at the user’s option, assuring the proper temperature during entire treatment.

Surgical Case Cart Washer Utensil Washer INSULATED CONSTRUCTION
Washer shall be insulated with 2-inch thick rigid fiberglass insulation.

Surgical Case Cart Washer Utensil Washer PERSONNEL SAFETY FEATURES
Emergency push/pull stop buttons are provided at all external doors to immediately terminate all processing functions, open all doors and activate all visual and audible alarms.
Two (2) red safety cables shall be installed inside the washer/dryer chamber (one on each side). If either cable is pulled, operation of the washer shall immediately stop by means of a mechanical safety switch system.  The unit must be reset by cycling the control power off and on, resetting the mechanical switch and depressing the “Cycle Start” button to resume operation.
All doors must be fully closed before a cycle can be initiated.  Once all doors are closed, the starting of any cycle is delayed for a period of ten (10) seconds during which all visual and audible alarms are activated.  All doors may be manually opened from within the chamber.  Opening of any door after activation of the cycle start button immediately terminates all treatment functions and activates all emergency alarms.  The doors are equipped to detect an obstruction during closing and will automatically open and activate all emergency alarms until the obstruction is removed and the controls have been reset.  The doors also are equipped with physical restraints to mechanically lock and restrict the doors from closing when the restraints are engaged.

Surgical Case Cart Washer Utensil Washer INCOMING UTILITY GAUGES
The washer shall be provided with incoming steam and hot water pressure gauges.

Surgical Case Cart Washer Utensil Washer AUTOMATIC FOUR PHASE TREATMENT CYCLE
The standard treatment cycle consists of a Detergent Wash, Non-Re-circulated Rinse, Vapor Removal and Hot Air Dry.  All cycle phases may be selected or deselected by the user. The cycle, once activated is completely automatic.

Surgical Case Cart Washer Utensil Washer
MULTICYCLE COLOR/GRAPHIC MICROCOMPUTER CONTROL SYSTEM
A solid-state microcomputer control system monitors and automatically controls all process operations and functions.  A twelve-cycle treatment menu may be programmed and retained to permit operating personnel to accommodate a wide variety of loads and processing requirements.  Individual cycle phase times and temperatures, and other key process parameters are programmable.  Each cycle may be locked in individually by supervision to insure process integrity.  Cycle programming may be controlled by a supervisory access code.   If the Strip Chart Printer option is purchased each cycle program may be reviewed and printed on demand.

Surgical Case Cart Washer Utensil Washer WATER TEMPERATURE BOOSTER
3.0.1 The washer is equipped with an instantaneous steam to water heat exchanger to raise the house hot water supply temperature from 120ºF to 190ºF.  The booster is inter-piped and inter-wired for automatic operation to supply heated water for the final rinse treatment phase of the cycle.

Surgical Case Cart Washer Utensil Washer RE-CIRCULATED HOT AIR DRYING SYSTEM
The carts are exposed to a re-circulated hot air drying system to assist in the drying of the exposed cart surfaces.  The hot re-circulated air is generated by
re-circulating air through a steam coil and blower arrangement capable of achieving a chamber temperature of up to 275º F at 80 PSI steam pressure.
An adjustable manual damper is provided in the blower intake plenum to provide
fresh make-up air to replace exhausted moist air.

Surgical Case Cart Washer Utensil Washer
OPERATION

At the beginning of the processing period, the operator activates the unit for automatic operation by pressing the power control switch, selecting the desired cycle and depresses the “Cycle Start” button. The safety feature sequence is activated, the load, unload and center doors close, the wash/rinse chamber sump fill and heat, and the hot air dryer is activated. The unit will then automatically advance to the ready state when all of the required operational conditions have been met.
When the ready state has been obtained, carts placed in the load queuing position(s) will automatically be indexed by the conveyor into the wash/rinse chamber, into the drying chamber and into the unload position(s).  The presence of each cart is monitored and only the specific chamber doors are activated and opened to advance the existing carts to the next position.  If the final unload position is occupied, the unit doors will remain closed and the carts will not be indexed through or out of the unit until the unload position is clear to accept additional carts.
The operator positions the first cart to be processed into the first load position.  The closed cart types must be placed into position with their doors latched open so the carts open side faces the tow side of the chamber.  Additional carts may be placed in the automatic load position without activation of the cycle start switch.  Personnel are only required to move carts into and take carts away from the load and unload zones.

At the end of the processing period, the operator places the unit in the standby mode by pressing the Standby Keypad.  All unit processing functions will immediately stop.  The wash/rinse chamber sump will automatically drain, the dryer blowers will stop and the load and center doors will open.  The unload door remains closed during the standby period. 

Surgical Case Cart Washer Utensil Washer Sequence of TREATMENTS

Surgical Case Cart Washer Utensil Washer DETERGENT WASH 
      Hot water from the house supply shall fill the sump to the required level and detergent is automatically added to the desired concentration. The treatment solution is re-circulated under pump pressure through the oscillating jet system onto the load. The wash solution is maintained in the sump for the following cycle.  The wash solution is periodically automatically pumped to drain after completion of a programmed number of wash treatment cycles.  Wash phase time and temperature is user programmable from 0 – 9999 seconds in 1-second intervals and 120 - 190?F in 1?intervals respectively.  The wash phase is operational only when a cart is in position within the wash/rinse section.

Surgical Case Cart Washer Utensil Washer RINSE
Hot water from house supply flows through the hot water temperature booster and is heated to 190 ºF. A rinse agent can be simultaneously added while the solution is sprayed at house supply pressure through the independent rinse spray system on the load. The rinse solution is collected in the sump and overflowed to drain. Rinse phase time is adjustable from 0 – 9999 seconds in 1- second intervals and temperature is adjustable from 120 - 190?F in 1? intervals. The rinse phase is only operational when a cart is in position within the wash/rinse chamber.

Surgical Case Cart Washer Utensil Washer  VAPOR REMOVAL
The unit stands idle for a sufficient period of time for the facility vent system to remove vapor from within the wash/rinse chamber.  Phase time is adjustable from 0 – 9999 seconds in 1-second intervals.

Surgical Case Cart Washer Utensil Washer  DRYING PHASE
Once the cart has advanced into the dryer chamber, heated air is circulated           through the chamber to dry all exposed surfaces of the cart.  The dry time is not    programmable and is a function of the total of the wash, rinse and vapor removal phase time settings. The re-circulating air temperature is user adjustable between 180ºF and 250ºF depending on steam supply pressure.  The dryer is operable during every cycle phase and is not dependent on the presence of a cart in the drying chamber.

Surgical Case Cart Washer Utensil Washer
CONSTRUCTION

The Surgical Case Cart Washer Utensil Washer base, wash chamber and chamber sump shall be of stainless steel and are of smooth construction without crevices and ledges for the potential build up of debris and contamination. The base and sump shall be of one-piece welded construction with the base containing integral door gutters and supports to accommodate the floor grating.
The Surgical Case Cart Washer Utensil Washer load, unload and center doors shall be of double wall construction, insulated with 1 inch thick rigid fiberglass insulation, and shall be mechanically powered, horizontal sliding, telescoping doors which operate as required to permit cart advance.  A Surgical Case Cart Washer Utensil Washer 12” x 16” Tempered Glass Observation Window that is provided in each of the load and unload doors.  All doors are immediately opened upon activation of any emergency stop switch or cable and can be manually opened from within the chamber.
Surgical Case Cart Washer Utensil Washer shall be insulated with 2-inch thick rigid fiberglass insulation covered by protective stainless steel panels.
Surgical Case Cart Washer Utensil Washer shall be Wash chamber floor shall consist of heavy-duty stainless steel grating covering the entire load area. Grating sections shall be easily removable for cleaning or maintenance.
The Surgical Case Cart Washer Utensil Washer chamber sump shall be equipped with a level control switch, automatic temperature control, automatic water fill port, overflow piping and a stainless steel steam heating coil for the heating of re-circulated treatment solutions.  The control system shall display and monitor re-circulated solution temperatures. 
The stainless steel steam coil shall heat the re-circulated wash solution shall be complete with a condensate return and steam traps. Steam coils shall be designed to ASME Section VIII, Div. 1, Unfired Pressurized Vessel Code, and shall be easily removable for cleaning or maintenance.
All Surgical Case Cart Washer Utensil Washer re-circulated treatments shall be under pump pressure.  The pump system shall be equipped with a direct reading, liquid filled pressure gauge.
The Surgical Case Cart Washer Utensil Washer shall be equipped with a transformer for the control circuit, magnetic starters with overload protection for all motors and all other electrical components required for the washer operation.
An exterior mounted light shall be provided to illuminate the wash chamber.
The Surgical Case Cart Washer Utensil Washer shall be equipped with pneumatically actuated ball valves to control the output of the pump to the jet system, drain or detergent solution return system.
The Surgical Case Cart Washer Utensil Washer shall be inter-piped and inter-wired so that only one connection is required for each service or utility.

Surgical Case Cart Washer Utensil Washer shall be supplied with three (3) 3/4” NPT detergent injection ports and dry electrical contacts for installation of automatic detergent injection pumps.  Washer sump shall also be equipped with two (2) 1" NPT half couplings for the connection of external devices.

Surgical Case Cart Washer Utensil Washer Programming will be in standard Military time and Fahrenheit temperature. Times and temperatures shall be expressed in seconds and full degree increments respectively.  An internal battery shall back up all cycle memory for up to ten years and permit completion of a cycle upon restoration of power following a power disruption.

TheSurgical Case Cart Washer Utensil Washer will be equipped with a highly visible Color Touch Controller Screen, displays cycle program data on demand and real time in-process cycle performance.  All cycle deviations are alarmed both visually and audibly and must be acknowledged by the operator.
Washer shall be provided with one (1) additional dry electrical contact for control for external damper(s) installed in the facility HVAC system (by customer).

Surgical Case Cart Washer Utensil Washer
MACHINE Configuration (select one):

Model 480LX RR, Right Hand Service Access, Right Hand Conveyor
Model 480LX LR, Left Hand Service Access, Right Hand Conveyor
Model 480LX RL, Right Hand Service Access, Left Hand Conveyor
Model 480LX LL, Left Hand Service Access, Left Hand Conveyor
Model 486LX RR, Right Hand Service Access, Right Hand Conveyor
Model 486LX LR, Left Hand Service Access, Right Hand Conveyor
Model 486LX RL, Right Hand Service Access, Left Hand Conveyor
Model 486LX LL, Left Hand Service Access, Left Hand Conveyor

Surgical Case Cart Washer Utensil Washer
Voltage (select one):

208 Vac. Three Phase 60 Hz 4 Wire, Three Phases & Neutral
208 Vac. Three Phase 60 Hz, Three Phases Only
230 Vac. Three Phase 60 Hz
460 Vac. Three Phase 60 Hz

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURES

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE INTEGRAL DATA COLLECTION SYSTEM
When the printer option is selected, chronological performance data may be stored within the unit and downloaded to the printer or a computer file on demand.  While capacity varies with cycle profile and the number of cycles run each day the typical capacity is 90 calendar days.  At the end of that period the unit will begin to overwrite the oldest stored data first.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE RS232 OR RS485 DATA COMPUTER PORT
Either an RS232 or RS485 port is provided to permit the capability to download cycle real-time performance data to a remote computer terminal.  Custom software may be provided to accommodate specific applications.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE MODEM
A modem is supplied for online service diagnostics, evaluations, and software upgrades. This allows factory-engineering personnel immediate access to the unit’s control system. Factory based service personnel will be able to troubleshoot, identify system malfunctions, and recommend repairs to local maintenance personnel within minutes of a malfunction reducing expensive onsite repair technician costs, allowing the customer to perform maintenance repairs internally, and minimize lost productivity due to possible equipment downtime. 
NOTE: This feature requires a dedicated analog phone line to be provided by the customer.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE FLOOR SPRAY HEADER
A stationary spray header is mounted in the floor of the unit to ensure complete exposure of the under side of the load to the treatment solution.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Drain Discharge Cool Down System with Cold Water Injection
The washer shall be provided with a cold-water inlet valve, integral with the drain line, to add cold water during draining for assistance in reducing the effluent temperature.  A temperature gauge shall be included on the cold-water inlet piping.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Drain Discharge Cool Down System with SEPARATE Tank
The washer shall be provided with a cool down tank to accept all pumped drain discharges.  Higher temperature effluent is cooled as it is mixed with cold water and gravity drained.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Drain Discharge Cool Down System with SEPARATE Tank and Temperature Assurance
The washer shall be provided with a cool down tank to accept all pumped drain discharges.  By mixing with cold water, discharges shall be cooled to below 140?F before gravity draining to building drain system.  The system shall be controlled through an automatically actuated ball valve and temperature probe, and shall not drain unless the effluent has been cooled to the proper temperature.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Exhaust Fan
Washer shall be provided with a stainless steel fan inter-wired with the microprocessor control system to exhaust residual vapors from within the wash chamber.  Fan shall be supplied complete with 3 phase, 60 Hz motor and magnetic starter for overload protection.  Fan shall be designed to push exhaust a short distance.  Fan shall be provided with remote-mounted grease fittings for convenient access.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE 90 Degree Automatic Damper
Washer shall be provided with an automatically actuated damper, mounted in the exhaust line and coordinated with the automatic cycle.  Damper shall be fully open during exhaust phase and partially open during washer operation.  Damper shall be designed with 90? angle for connection to horizontal exhaust vent system.
Stainless Steel Service Access Panels
Stainless steel side panels and supports shall be provided to enclose the service the component side of the washer.  Panels shall be hinged for easy access.
Split Base for Special Entry into Building
Base shall be constructed of two (2) parts and welded together at the facility during installation.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Surgical Case Cart Washer Utensil Washer SEISMIC TIE DOWN
Washer shall be designed to comply with Seismic Zone 3 & 4 requirements.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Alkaline Detergent Injection System
Washer shall be equipped with a liquid detergent injection pump, detergent pick-up tube, 50 feet tubing, conductivity controller and a conductivity probe.  The unit automatically monitors and directly injects alkaline detergent into the chamber sump during the alkaline wash phase.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Low Pressure Steam Piping
The washer shall be provided with special steam line piping to accommodate 10 to 30 psi steam pressure.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE High Altitude Treatment Pump
Washer shall be provided with a high altitude treatment pump for installations where the elevation is at least 3000 feet above sea level.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE pH Neutralization and Detergent Monitoring System
Washer shall be equipped with control hardware and pH probes to monitor and control the pH level of drain discharge.  Each time the washer attempts to drain, the pH level shall be checked.  If the pH level is within a pre-set range, the washer shall drain.  If not, the proper neutralizing agent shall be injected and the solution re-circulated and tested again.  This process shall be repeated three times until all parameters are met.  If after the third test, the parameters are not met, an alarm shall sound.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE AIR COMPRESSOR
The washer is provided with an air compressor inter-wired and inter-piped with the automatic control system to supply the air demand required by the washer.  The compressor will be supplied with regulator/filter/oiler and a holding tank.

Surgical Case Cart Washer Utensil Washer
OPTIONAL FEATURE Remote Mounted Control Column
Washer shall be provided with an additional 20 feet of conduit and connections for remote mounting of the control column.  Brackets to mount the control column shall be provided by customer.

Surgical Case Cart Washer Utensil Washer
ACCESSORIES

Barrier Wall Flange Assembly
Washer shall be provided with a stainless steel trim flange to enclose the opening between one end of the washer and the masonry wall opening.
Recessed one (1) wall
Recessed two (2) walls

Surgical Case Cart Washer Utensil Washer
Utensil Cart
A Stainless steel cart is provided for processing utensils and other items used in
a central supply department.  The rack is transported by casters with roller bearings
of stainless steel construction.

Surgical Case Cart Washer Utensil Washer
MATERIALS OF CONSTRUCTION
Standard Items
Material   1.Base & Chamber Sump 12 Gauge 304 SS
Wash Solution Reservoir(s) 12 Gauge 304 SS #3 Fin.
Door Panels 16 Gauge 304 SS #3 Fin.
Side and Top Panels 14 Gauge 304 SS #3 Fin.
Re-circulating Valves and Components 304 SS
Water and Steam Re-circulating Piping- Internal 304 SS
Drain Piping 304 SS
Drain Valves & Components 304 SS
Steam Piping - External Schedule 80 Black Iron
Water Piping - External Brass
Spray Jets 304 SS
Steam Coils 304 SS
Tow Conveyor 304 SS

Surgical Case Cart Washer Utensil Washer
MATERIALS OF CONSTRUCTION
Optional Items
Barrier Flange 20 Gauge 304 SS #3 Fin.

Surgical Case Cart Washer Utensil Washer
UTILITY DATA

EElectrical      3-60-208/230/460 V25 Hp
S1Steam2 ½”NPT    30-80 PSI1000 Lbs/Hr   Flow Rate - Washer
S2Steam2 ½”NPT    30-80 PSI1400 Lbs/Hr   Flow Rate- Dryer
HWHot Water1 ½” NPT     35 PSI    70 Gal/Load Consumption
50 GPM Flow Rate 140ºF Minimum ± 5º F
CCondensate 1” NPT - 2 Connections                              
DDrain 3” Floor Drain-Minimum         120 GPM Flow Rate @ 200?F
V1Vent12” ID Duct   700 CFM -Washer
Saturated Vapor @ 190 ?F

V2Vent12” ID Duct   700 CFM -Dryer
Saturated Vapor @ 190 ?F
CW    Cold Water      1 ½” NPT         35   PSI   60 GPM @ 50?F± 5º F
AAir    ½” NPT        100 PSI        6 CFM Flow Rate

Surgical Case Cart Washer Utensil Washer
Surgical Case Cart Washer Specifications

Surgical Case Cart Washer 480LX
Surgical Case Cart Washer 486LX
Contact us for your 'competitive' pricing.
Surgical Case Cart Washer Detergent and
Surgical Instrument Cleaners
Surgical Case Cart Washers

Automated... 
Surgical Case Cart Washer.

Manual...  feed
Surgical Case Cart Washer.
480 LX 486 LX  Surgical Case Cart Washer Utensil Washers Surgical Closed Case Cart Washers Utensil Washers. Surgical Closed Case Cart Washers.
480 LX 486 LX  Surgical Case Cart Washer Utensil Washers Surgical Closed Case Cart Washers Utensil Washers. Surgical Closed Case Cart Washers.
Surgical Case Cart Washer
manufactured in the USA.

Surgical Case Cart Washer  nonproprietary
locally available replacement parts.